Installation

 

26 

MI 

5.4.3 Ambient Background Temperature Compensation 

The  sensor  is  capable  of  improving  the  accuracy  of  target 
temperature  measurements  by  taking  into  account  the  ambient  or 
background  temperature.  This  feature  is  useful  when  the  target 
emissivity  is  below  1.0  and  the  background  temperature  is 
significantly  hotter  than  the  target  temperature.  For  instance,  the 
higher  temperature  of  a  furnace  wall  could  lead  to  hotter 
temperatures being measured especially for low emissivity targets. 
Ambient background temperature compensation compensates for the 
impact  of  the  reflected  radiation  in  accordance  to  the  reflective 
behavior of the target. Due to the surface structure of the target, some 
amount of ambient radiation will be reflected and therefore added to 
the  thermal  radiation  that  is  collected  by  the  sensor.  The  ambient 
background  temperature  compensation  compensates  the  final  result 
by  subtracting  the  amount  of  ambient  radiation  measured  from  the 
sum of thermal radiation the sensor is exposed to. 

The  ambient  background  temperature  compensation 
should  always  be  activated  in  case  of  low  emissivity 
targets  measured  in  hot  environments  or  when  heat 
sources are near the target! 

Three 

possibilities 

for 

ambient 

background 

temperature 

compensation are available: 

• The  internal  sensor  head  temperature  is  utilized  for 

compensation 

assuming 

that 

the 

ambient 

background 

temperature  is  more  or  less  represented  by  the  internal  sensor 
head temperature. This is the default setting. 

• If  the  background  ambient  temperature  is  known  and  constant, 

the user may give the known ambient temperature as a constant 
temperature value