Maintenance

 

60 

MI 

9 Maintenance 

Our  sales  representatives  and  customer  service  are  always  at  your 
disposal  for  questions  regarding  application  assistance,  calibration, 
repair,  and  solutions  to  specific  problems.  Please  contact  your  local 
sales representative if you  need  assistance. In many cases, problems 
can be solved over the telephone. If you need to return equipment for 
servicing,  calibration,  or  repair,  please  contact  our  Service 
Department  before  shipping.  Phone  numbers  are  listed  at  the 
beginning of this document. 

9.1 Troubleshooting Minor Problems 

Symptom

Probable Cause

Solution

No output

No power to instrument

Check the power supply

Erroneous
temperature

Faulty sensor cable

Verify cable continuity

Erroneous
temperature

Field of view obstruction

Remove the obstruction

Erroneous
temperature

Window lens

Clean the lens

Erroneous
temperature

Wrong emissivity

Correct the setting

Temperature
fluctuates

Wrong signal processing

Correct Peak/Valley Hold or Average
settings

Temperature
fluctuates

No ground for the head

Check wiring / grounding

Table 6: Troubleshooting 