4 advanced peak hold, Operation, 42 mi – RayTek MI Miniature Infrared Sensor User Manual

Page 50

Advertising
background image

Operation

 

42 

MI 

6.4.4 Advanced Peak Hold 

This function searches the sensor signal for a local maximum (peak) 
and  writes  this  value  to  the  output  until  a  new  local  maximum  is 
found.  Before the algorithm restarts  searching  for a local maximum, 
the  object  temperature  has  to  drop  below  a  predefined  threshold.  If 
the  object  temperature  raises  above  the  held  value  which  has  been 
written  to  the  output  so  far,  the  output  signal  follows  the  object 
temperature  again.  If  the  algorithm  detects  a  local  maximum  while 
the  object  temperature  is  currently  below  the  predefined  threshold 
the  output  signal  jumps  to  the  new  maximum  temperature  of  this 
local maximum. Once the actual temperature has passed a maximum 
above  a  certain  magnitude,  a  new  local  maximum  is  found.  This 
magnitude is called hysteresis. 

 

Figure 28: Advanced Peak Hold 

The advanced peak hold function is only adjustable by means of the 
DataTemp MultiDrop Software. 

output temperature
object temperature

hysteresis

th

reshold

Temp

Time

Advertising