Programming guide mi 71 – RayTek MI Miniature Infrared Sensor User Manual

Page 79

Advertising
background image

Programming Guide 

MI 

71 

There are eight entries possible for emissivity setting (1) and a related 
set point (threshold) (2). To be able to write or read these values, use 
the following commands: 
EP=2 

set pointer for table entry, e.g. to line 2 (3) 

RV=0.600 

set the emissivity value for line 2 to 0.600 (4) 

SV=220.0 

set the set point (threshold) for line 2 to 220.0 (5) 

 

 

Figure 46: Table for Emissivity and Set Points 

To activate these emissivity settings, you need to have the 3 external 
inputs (FTC) connected. According to the digital combination on the 
FTC wires, one of the table entries will be activated, see section 5.4.2 

Emissivity Setting via Digital Selection

 on page 25. 