Programming Guide

 

74 

MI 

XF 

factory default values will be set 

11.6.5 Lock Mode 

The  access  to  the  unit  is  possible  via  serial  interface  (software)  and 
via  the  direct  user  input  (mode  buttons,  LCD  display).  It  is  possible 
to lock the buttons. This allows access the unit only via software. 

J=L 

direct user input via mode buttons denied 

Alternatively  the  unit  can  be  unlocked  by  pressing  the  <Mode/Up> 
buttons simultaneously for three seconds. 

11.6.6 Mode Setting for the Digital Input FTC3 

The  digital  input  FTC3  (see  section  5.4.4 

Trigger  and  Hold  Function

 

on page 28) can be used as follows: 

XN=T 

FTC3 as trigger 

XN=H 

FTC3 with hold function 

11.6.7 Changing the Sensing Head Calibration Data 

If  it  is  necessary  to  exchange  the  sensing  head,  you  must  set  the 
calibration data for the new sensing head: 

XZ=0123 4567 FFFF FFFF 

according to the head calibration data 

For MID/MIC‐models the last eight  numbers are not used and must 
be set to „F“ like shown in the example above. For MIH‐models and 
specials (like MTB or G5) all sixteen numbers are used. 

11.6.8 Ambient Background Temperature Compensation 

In  case  of  compensating  the  ambient  background  temperature,  the 
following modes are available: 

AC=0 

no compensation 

AC=1 

compensation  with  a  constant  temperature  value  set 
with command A. 