3 sensing head exchange, Maintenance mi 63 9.3 sensing head exchange – RayTek MI Miniature Infrared Sensor User Manual

Page 71

Advertising
background image

Maintenance 

MI 

63 

9.3 Sensing Head Exchange 

Sensing  heads  and  electronic  boxes  can  only  be 
interchanged in accordance to the following table! 

MID02 MIC02 MID10 MIC10 MIH10 MID20 MIC20 MIH20

MID02

x x x x

MIC02

x x x x

MID10

x x x x

MIC10

x x x x

MIH10

x

MID20

x

MIC20

x

MIH20

x

The  head  exchange  requires  to  type  in  the  new  sensing  head 
calibration data printed on the cable as follows: 

1.  To  exchange  the  sensing  head,  disconnect  the  power  of  the 

unit. 

2.  Connect  the  wires  for  the  new  sensing  head  according  to  the 

color description on the printed circuit board. 

3.  Switch the power for the unit to ON. 
4.  Press simultaneously the <Mode/Down/Up> buttons. 
5.  Four characters appear in the display (former values). Type in 

the  new  designator  (A)  using  the  <Down/Up>  buttons.  Press 
the <Mode> button. 

6.  The  second  block  of  four  characters  appears  in  the  display 

(former  values).  Type  in  the  new  designator  (B)  using  the 

Advertising