4 dealer calibration, Leaving intermediate calibration, User calibration on/off – Raymarine ST60 Tridata User Manual

Page 51: Response settings

Advertising
background image

Calibration

39

Leaving Intermediate calibration

Hold down the

depth

and

speed

keys for 2 seconds, to save your settings,

exit Intermediate calibration and resume normal operation.

4.4 Dealer calibration

The Dealer calibration procedures enable the following parameters to be
set:

User calibration on/off.

Speed response.

Depth response.

Boat show mode on/off.

Dealer calibration also gives access to the Factory defaults screen. This
enables you to re-apply the factory settings if you want to reset the
instrument to a known operating condition.

To commence Dealer calibration, hold down the

depth

and

speed

keys

together for approximately 12 seconds, to select the Dealer calibration
entry page (see Dealer calibration diagram). Then press the

trip

and

reset

keys together, to enter the calibration screen sequence.

Use the depth key to move from screen to screen and the trip or the rest
key to set the required values at each screen.

User calibration on/off

Press either the

trip

or

reset

key to toggle the User calibration on or off as

required. With off selected, User calibration and Intermediate calibration
are both disabled.

Response settings

The response values for both SPEED and DEPTH determine the frequency
at which information is updated. A low number provides a smooth
response and a high number a much livelier update.

Use the

trip

(decrement) and

reset

(increment) keys to set the required

value. Response values are from 1 to 15.

81040_1.book Page 39 Thursday, November 7, 2002 10:21 AM

Advertising