Table a- 3: waveform horizontal deflection – Tektronix 1740A/1750A User Manual

Page 106

Advertising
background image

Appendix A: Performance Specification

A-6

1740A/1750A Series Waveform/Vector Monitor User Manual

Table A- 3: Waveform Horizontal Deflection

CATEGORY

DESCRIPTION

Sweep

Req:

Synchronization: Sweep triggered by horizon-
tal and vertical sync pulses

RI:

Sweep Length:

≈12 divisions

RI:

Sweep freeruns without input

Sweep Timing Accuracy

Req:

1 Line:

5 s/division

±1%

2 Line:

10 s/division

±1%

RI:

1 Field: displays one full field, including
field rate sync

2 Field: displays two full fields and
the field rate sync between them

Sweep Linearity

Req:

1 line:

±1%

2 line:

±1%

Magnified Sweep Accuracy

Req:

1 Line: 0.2 s/division

±1%

2 Line: 1.0 s/division

±1%

Magnified Sweep Linearity

Req:

1 line:

±1%

2 line:

±1%

Horizontal Position Range

Req:

Any portion of the synchronized sweep can be
positioned on screen in all sweep modes.

External Horizontal Input

Req:

2 divisions/volt,

±2%

RI:

Menu is selected and enabled by RE-
MOTE connector ground closure.

Remote Sync

RI:

Input Amplitude: TTL level

RI:

Frequency: 25 Hz to 100 Hz positive
edge-triggered sweep

RI:

Enabling Signal: TTL low or ground
closure

Advertising