Filter Menu

(

WAVEFORM

Mode Only)

The

FILTER

menu is available for

WAVEFORM

mode only.

Note that the instrument must be in

WAVEFORM

mode to ac-

cess the R--Y displays. When the

FILTER

button is pushed in

WAVEFORM

mode, an on-screen menu allows the following

selections. The bezel buttons are self-cancelling.

H

FLAT —

provides flat (normal) response.

H

LUM —

provides a low-pass response to display the

luminance portion of the composite video signal.

H

CHROM —

provides bandpass response centered on

the chrominance subcarrier frequency, and displays
frequencies around the subcarrier frequency.

H

DIFF —

is a differentiated step filter (linearity steps

are translated into impulses for amplitude compari-
son).

H

R--Y —

selects demodulated chrominance-versus-

time display. The chrominance is demodulated on
the R--Y (V) axis when burst is lined up on the
normal axis. The

VECTOR PHASE

control can adjust

the demodulator phase to any axis.

H

SCH R--Y (1750A--Series Only) —

demodulates the

SCH sync-locked oscillator. This display is useful
for viewing SCH variations versus time.

H

F + L —

displays the video signal both flat and

luminance filtered. The display is in parade or
overlay mode, as selected through the on-screen
menu. When

PARADE

is selected, the flat display is

on the left.