With waveform display – Tektronix 1740A/1750A User Manual

Page 68

Advertising
background image

Using the Menus

1740A/1750A Series Waveform/Vector Monitor User Manual

3-13

If it becomes necessary to compensate for minor shifting of
markers due to drift or external magnetic fields (particularly
if the instrument is moved after the markers are set), enter
the

CRT

menu and select

READOUT

. Use the bezel controls to

adjust vertical position

(VPOS)

and horizontal position

(HPOS)

until the center of the displayed test pattern is at the

center of the graticule.

With Waveform Display

Voltage Cursors can be used to make accurate signal ampli-
tude measurements. With

VOLT

selected, use the left bezel

control to position the

VOLT1

cursor at the upper excursion of

signal to be measured, and the center bezel control to posi-
tion the

VOLT2

cursor at the lower excursion. The on-screen

readout gives the voltage difference between the two levels.

When

GAIN

is changed, the voltage cursors maintain their

relationship with the signal. For example, if the

VOLT2

cursor

were set to the signal sync tip in

X1 GAIN

, it would also be at

sync tip in

X5

gain

, X10

gain

,

or with

VAR

gain on.

Timing Cursors can be used to make accurate time measure-
ments. When

TIME

is selected, position the

TIME1

cursor at

the left excursion of the signal to be measured, and the

TIME2

cursor on the right. The on-screen readout gives the time
difference between the two points.

When

MAG

is selected, the timing cursors maintain their rela-

tionship with the signal. For example, if the

VOLT1

cursor

were set to the rising edge of sync, it would also be set there
with

MAG

selected.

Markers. When

MARK

is selected, use the bezel controls to

vertically position the markers where desired with respect to
the graticule markings. These markers are designed to be
used for reference points; there is no amplitude readout. The
markers are not affected by

GAIN

changes; if a marker were

set to

1

00 IRE

in

X1 GAIN

, it would still be set to

100 IRE

with

the

MAG

on.

Advertising