Offset, Calibrate – Tektronix 1740A/1750A User Manual

Page 76

Advertising
background image

Using the Menus

1740A/1750A Series Waveform/Vector Monitor User Manual

3-21

OFFSET

H The

VERTICAL POSITION

control normally positions

all channels together. When

OFFSET

is enabled, the

bezel controls can be used to offset input channels 2
(

A

or

B

) and 3 (

A

or

B

) with respect to channel 1 (

A

or

B

).

H While the offsets are enabled, the

VERTICAL POSI-

TION

control will still position all channels together,

with the offset amount remaining constant. When

OFFSET

is turned off, the input channels will be

re-aligned. The selected offset amount will be
recalled the next time

OFFSET

is enabled.

H When

A2A3 OFFSET

is on, the center and right bezel

controls are assigned to

CH A2

and

CH A3

offsets.

On-screen labels appear just above each control.
Turning these controls will offset the

CH A2

and

CH

A3

signals with respect to

CH A1

.

When

B2B3

offset

is on, the same two controls are assigned to

CH B2

and

CH B3.

H When both

A2A3

and

B2B3

offsets are on, the user

assigns the controls to either

A

or

B

offsets. The

CONTROL

selection only appears when both

A

and

B

OFFSETS

are selected.

H An offset will remain in effect until the user turns

the offset

OFF.

(Exiting the

CONFIG

menu will not

affect the

OFFSET ON / OFF

setting.)

CALIBRATE

These choices are offered when

CALIBRATE

is selected:

H

CAL SIG

Turns the 1 V internal calibrator signal

on/off.

H

GAIN CAL

Turns on the gain calibration adjust-

ment on/off.

Advertising