Calibration

4-18

1740A/1750A Series Waveform/Vector Monitor User Manual

Set Horizontal Gain

1.

To set horizontal gain, push the

CONFIG

menu

button. Select

CALIBRATE

, then

CAL SIG ON

and

100

KHZ

.

Select

GAIN CAL ON

.

a.

In

1 LINE SWEEP

, adjust the

HORIZ CAL

control for one-half cycle per major grati-
cule division; in

2 LINE SWEEP

, adjust for

one full cycle per major division.

b.

Pushing

RESET HCAL

will cancel any gain

adjustment and reinstate the factory gain
setting.

Set Vector Gain

1.

To adjust vector gain, select

VECTOR

display mode

and push the

CONFIG

menu button. Select

CALI-

BRATE

, then

CAL SIG ON

and

GAIN CAL ON

.

a.

Adjust

VECT CAL

so that the calibration

signal circle falls on the compass rose
(outer ring).

b.

Pushing

RESET VEC CAL

will cancel any

gain adjustment and reinstate the factory
gain setting.