Table a- 4: measurement cursors, Table a- 5: rgb/yrgb – Tektronix 1740A/1750A User Manual

Page 107

Advertising
background image

Appendix A: Performance Specification

A-7

1740A/1750A Series Waveform/Vector Monitor User Manual

Table A- 4: Measurement Cursors

CATEGORY

DESCRIPTION

Waveform Accuracy

Req:

Voltage: 0.5%
Timing: 0.5%, for line rate sweeps

RI:

Typically 0.5% for field rate sweeps

Vector Accuracy

Req:

Gain:

±1.5%

Phase:

±1°

RI:

Measured with respect to the Color
Bar signal

Table A- 5: RGB/YRGB

CATEGORY

DESCRIPTION

RGB/YRGB

Req:

Staircase input gain: 0.8 V/division

±10%

RI:

RGB Sweep Length:
1 Field

≈30% of normal

1 Line

≈ 30% of normal

RI:

YRGB Sweep Length:
1 Field

≈ 25% of normal

1 Line

≈ 25% of normal

RI:

Maximum staircase operating signal: DC
signal plus peakacnotto exceed --12 V
to+12 V. Line or field rate sweep.

RI:

Peak-to-peak acsignal not to exceed

12 V.

RI:

Sweep Repetition Rate: Field or line rate
of displayed video or external
sync signal as selected by the front-
panel sweep selection.

Advertising