Location, Mounting, and Ventilation

Release 1.1

29

Location, Mounting, and Ventilation

Use the power supply in rack-mounted applications only. The power supply is
designed to fit in a standard 19 in. (483mm) equipment rack.

Rack

Mounting

To install the power supply in an equipment rack:

1. Open the box containing the unit. See

Figure 2.2

.

2. With the help of at least one other person, lift the unit out of its package and slide

it into an empty space in a mounting rack equipped with rails that are rated to
support the unit’s weight. See

Figure 2.3

.

3. While your assistant(s) holds the unit steady, fasten it to the rack by inserting

bolts through the mounting brackets on either side of the front panel and securing
them with a washer and nut.

4. The front panel mounting brackets are designed to prevent the unit from sliding

out of the rack, not to support its full weight. Provide adequate support for the
rear of the unit without obstructing the ventilation inlets. Use slide rails as
illustrated in

Figure 2.3

or slide brackets attached to the 8-32 mounting holes on

each side of the unit. Follow the manufacturer’s instructions to install rails or
slides.

WARNING- High Energy and High Voltage
Ensure that the 8-32 rack mounting screws do not extend more than 1/8 in. (3.0mm)
into the sides of the power supply.

!

CAUTION
The power supply is too heavy for one person to safely lift and mount. To avoid injury,
ask a co-worker for assistance.