Ƽ warning – Activeforever Invacare Bariatric Electric Hospital Bed (Bariatric Home Care Hospital Bed) User Manual

Page 31

Advertising
background image

SECTION 5—PACKAGING, HANDLING AND SET-UP

Part No 1123842

31

IVC™ Bariatric Bed

8. Release the Hi/Lo motor output shaft spring-loaded coupler by pushing in against the

Hi/Lo motor output shaft spring-loaded coupler and turning clockwise.

NOTE: The coupling will release and engage the foot end top gear box output shaft cross pins.

9. Ensure that all components are securely installed.

ƽ WARNING

Close the drive shaft covers over all drive shafts that are not in use before placing

the bed in service. Otherwise, damage or injury may occur.

FIGURE 5.8 Assembling/Installing/Removing Drive Shaft

FOOT END

HEAD END

Bed End

Hi/Lo Motor Output Shaft

Spring-Loaded Coupler

Top Gear Box Output

Drive Shaft

DETAIL “A”

Hi/Lo Motor Output Shaft

(Full Electric Only)

Hi/Lo Motor Output Shaft

Spring-Loaded Coupler

Locked Position

Drive Shaft

Unlocked Position

Drive Shaft

Hi/Lo Motor Output Shaft

(Full Electric Only)

Hi/Lo Motor Output Shaft

Spring-Loaded Coupler

Hi/Lo Motor

Top Gear Box Output

Drive Shaft

Bottom Gear Box

Output Drive Shaft

(Covered)

Output Side

Drive Shaft Cover

Gear Box

Input Side

Input Drive Shaft

(Covered)

Gear Box

Advertising