Section 7- troubleshooting, Troubleshooting: electrical, Section 7—troubleshooting – Activeforever Invacare Bariatric Electric Hospital Bed (Bariatric Home Care Hospital Bed) User Manual
Page 35: Ƽ warning
SECTION 7—TROUBLESHOOTING
35
SECTION 7—TROUBLESHOOTING
ƽ WARNING
Read all general guidelines, warnings and cautions before performing any
repair/maintenance on the bed. All repair and maintenance on the bed should be
performed by a qualified technician.
Troubleshooting: Electrical
SYMPTOM
PROBABLE CAUSE
SOLUTIONS
Bed section does not
move.
End of stroke reached.
Operate opposite button.
Full Electric Bed: Hi/Lo
function does not work.
Bed not plugged in.
Ensure power cord is plugged into power
source.
Pendant not functioning.
Ensure Pendant correctly connected to
junction box. Refer to FIGURE 5.9.
Ensure Pendant cable is not entangled/pinched.
Cables
Ensure motor cable(s) correctly connected.
Refer to FIGURE 5.9.
Ensure cables are not entangled/pinched.
Call the dealer for repair/replacement of
cables and/or motors.
Household fuse or breaker blown.
Check household fuse/breaker box.
Motors are not functioning.
Call dealer.
Hi/Lo Motor
Ensure Hi/Lo motor cable is correctly
connected. Refer to FIGURE 5.9.
Ensure cables are not entangled or pinched.
Ensure drive shaft is correctly installed.
Bed Spring does not stop
moving.
Pendant
Check for depressed/stuck buttons. If button is
stuck, call dealer for repair.
Bed Hi/Lo movement
does not stop.
Check cables. If damaged, call dealer.