SECTION 7—TROUBLESHOOTING

IVC™ Bariatric Bed

36

Part No 1123842

Troubleshooting: Mechanical

SYMPTOM

PROBABLE CAUSE

SOLUTIONS

Bed does not stay in place. Locking casters not locked.

Lock casters. Refer to Installing Casters on
page 26.

Bed sections do not move. Pull tube

Ensure pull tubes are properly connected.
Refer to FIGURE 5.9.
Check pull tube(s) for binding. Call dealer for
repairs.

Bed ends move in opposite
directions (electrical or
manual bed).

Drive shaft connected to wrong
gear box shaft.

Reinstall drive shaft. Refer to Assem-
bling/Installing/Removing Drive Shaft on
page 30
.

Hi/Lo function does not
work.

Drive shaft disconnected.

Reinstall drive shaft. Refer to Assem-
bling/Installing/Removing Drive Shaft on
page 30
.

Pressing function on
pendant results in motion
other than that intended.

Motor cable connections to
junction box incorrect.

Verify connections. Refer to FIGURE 5.9.