4 Mounting

VEGASOURCE 35 •

38132-EN-130415

Mounting brackets, angled attachments as well as mounting clips are

available for mounting.
Vertical pipeline, 30° inclined, diameter 50 … 100 mm

(1.97 … 3.94 in)

For pipeline diameters 50 … 100 mm (1.97 … 3.94 in), a diagonal

radiation path is recommended. The distance of the beam through the

medium is thus longer and an improved measuring effect is achieved.

For this, the optional lead shielding for the detector is recommended

in order to avoid influence from secondary radiation sources.

Fig. 13: 30° measurement arrangement on a pipeline with diameter

50 … 100 mm (1.97 … 3.94 in)

Vertical pipeline, diameter 50 … 420 mm (1.97 … 16.54 in)

For pipeline diameters 50 … 420 mm (1.97 … 16.54 in), a straight ra-

diation path is possible. The radiation-based sensor can be mounted

either horizontally or vertically.