28

5 Connection of the pneumatic switchich facility

VEGASOURCE 35 •

38132-EN-130415

5 Connection of the pneumatic switchich

facility

5.1 Connection of the proximity switches

These instructions apply to source containers VEGASOURCE 35 ver-

sion K, L, M, N with pneumatic switching facility.

Always keep in mind the following safety instructions:

Connect only in the complete absence of line voltage

If overvoltage surges are expected, overvoltage arresters should

be installed

Connect the proximity switches according to the following diagrams.

Take note of the general installation regulations. As a rule, connect

VEGASOURCE 35 to vessel ground (PA), or in case of plastic ves-

sels, to the next ground potential. On the upper side of the instrument

protective cover there is a ground terminal between the cable entries.

This connection serves to drain off electrostatic charges. In Ex ap-

plications, the installation regulations for hazardous areas must be

given priority.
The data for power supply are specified in chapter "Technical data".

The instrument is connected with standard two-wire cable without

screen.
Use cable with round cross-section. A cable outer diameter of

5 … 10 mm (0.2 … 0.39 in) ensures the seal effect of the cable gland.

If you are using cable with a different diameter or cross-section,

exchange the seal or use a suitable cable gland.

Suitable proximity switch: Pepperl+Fuchs 181094-NCB2-12GM35-

NO-10M

1

2

1 2 3 4

br bl br bl

+ - + -

Fig. 21: Connection of the proximity switches
1 Proximity switch for switch position EIN - ON (terminals 1 and 2)

2 Proximity switch for switch position AUS - OFF (terminals 3 and 4)

Safety instructions

Potential equalisation

Connection cable

Electrical connection