7 Maintenance and fault rectification

VEGASOURCE 35 •

38132-EN-130415

7.3 Testing the switching facility

Test the function of the switching facility (source holder) in regular

intervals.
Measurement of the local dose rate

1. Loosen the fixing pin (version B) or remove the lock (if present) as

described in chapter "Setup".

2. Move the source insert as described in chapter "Setup" several

times from the ON to the OFF position and vice versa. The source

insert should be easily movable and must have no traces of corro-

sion in the visible area.
If the source insert cannot be moved from ON to OFF position,

then you have to follow the instructions in paragraph "Reaction in

an emergency case".
If it is hard to move the source insert or there are other traces

of a malfunction, then the source insert mus be locked in the

position OFF and the responsible radiation safety officer must be

informed.
In case of corrosion: Follow the instructions in chapter "Mainte-

nance/Measures in case of corrosion".

1. Remove the padlock (see chapter "Setup")
2. Switch the source insert by means of compressed air from the po-

sition OFF to position ON. The source insert should move without

interruption to position "ON".

Caution:

Do not grasp into the indication window of the indication plate as long

as the pneumatics switches over.
3. Reduce the pressure to below 2.5 bar (36 psi). The source insert

must move back to position OFF.
If the source insert does not move continuously or shows signsl

of a possible malfunction, then the source insert must be locked

in position OFF and the responsible radiation safety officer must

be informed.
If the source insert cannot be moved from ON to OFF position,

then you have to follow the instructions in paragraph "Reaction in

an emergency case".
In case of corrosion: Follow the instructions in section "Mainte-

nance/Measures in case of corrosion".

7.4 Tightness test

The tightness of the source capsule must be checked in regular inter-

vals. The sequence of the tightness test (wipe test) must correspond

to the specifications of the authorities or the handling permit.
Note:

A tightness test is not only required as a regular test but must be car-

ried out after each incident that could impair the cover of the source.

In such case, the tightness test must be prescribed by the responsible

radiation safety officer under consideration of the necessary regua-

Source holder with

manual switching facility

Source holder with pneu-

matic switching facility