6 Set up with the display and adjustment module

MINITRAC 31 • Profibus PA

43836-EN-131119

The natural radiation on earth influences the accuracy of the meas-

urement.
With this menu item the natural background radiation can be faded

out.
For this purpose, the MINITRAC 31 measures the natural background

radiation and sets the pulse rate to zero.
In the future, the pulse rate from this background radiation will be

automatically deducted from the total pulse rate. This means: only

the component of the pulse rate originating from the source will be

displayed.
The source container must be closed for this setting.

In this menu item you can select the units of the process value and

the temperature.

In this menu item you can enter the measuring range (min. and max.

process value) of the sensor.
These settings influence the current output of the sensor.
Enter in the menu window "Max. process value" the max. level (full),

for example in "m". This corresponds to an output current of 20 mA.
Enter in the menu window "Min. process value" the min. level (empty),

for example in "m". This corresponds to an output current of 4 mA.

In this menu item you can carry out the adjustment of the sensor.
Due to the measuring principle, there is no linear relationship between

pulse rate and level. Hence, this adjustment (i.e. linearisation) must in

any case be carried out.
Note:

If you cannot fill the vessel with the original medium, it is also possible

to carry out the adjustment with water.
Prerequisites:
Radiation is switched on - Source container is set to "ON"
The vessel is either completely filled (100 %) or completely emptied

(0 %).
Depending on the fact if the vessel is filled or emptied, you can car-

ried out first the full or the empty adjustment. The MINITRAC 31 sorts

the points automatically according to their level.
Select "Show table" to display and edit the linearization points.
Select "Linearization - New" to enter the first point.
Select "Determine count rate" to enter the first point.
The determination of the actual count rate lasts 2 minutes. After the

count rate has been determined, you can accept the value (ct/s).
Enter now the corresponding level (m).
By doing so, you assign a respective level to the actual count rate.

Setup - Background

radiation

Setup - Units

Setup - Adjustment

Setup - Linearization