6 Set up with the display and adjustment module

MINITRAC 31 • Profibus PA

43836-EN-131119

In menu item "Scaling" you define the scaling format on the display

and the scaling of the measured values for 0 % and 100 %.

Measured value min.

Measured value max.

To damp process-dependent measured value fluctuations, you can

set a time of 0 … 999 s in this menu item.

The default setting is a damping of 0 s.

In this menu item you can activate the relay output and determine its

function as well as the switching points.
When the output of the process values is set, you can choose be-

tween overfill and dry run protection.
The relay outputs of the sensor react accordingly.
You can choose "no" reference value. In this case, the relay output

operates as fail safe relay.
Caution:

Independent of the selected reference value, the relay will deenergize

in case of failure.
Additional adjustments

When a reset is carried out, all settings (with only a few exceptions)

are reset. The exceptions are: PIN, language and SIL.

Setup - AI FB1 - Scaling

Setup - AI FB1 - Damping

Setup - Relay

Additional adjustments

- Reset