VEGA MINITRAC 31 Profibus PA User Manual

Page 50

Advertising
background image

50

6 Set up with the display and adjustment module

MINITRAC 31 • Profibus PA

43836-EN-131119

Peak values of temperature: Resetting of the measured min. and

max. temperatures to the actual measured value.
The following table shows the default values of the instrument. The

values apply for the application "Level". First of all you have to select

the application.
Depending on the instrument version, not all menu items may be

available or they may be differently assigned:

Menu

Menu item

Default value

Setup

Sensor address

126

Measurement loop name

Sensor

Isotope

Cs-137

Application

Density

Background radiation

0 ct/s

Unit of the process value

g/cm

3

Temperature unit

° C

Adjustment min.

Min. process value = 0.5 g/

cm

3

Adjustment max.

Max. process value = 1.5 g/

cm

3

Inner diameter

0.2 m

Linearization

90000 ct/s = 0,5 g/cm

3

AI FB1 Channel

Process value

AI FB1 scaling unit

%

AI FB1 scaling

0 g/cm

3

= 0 %

99.999 g/cm

3

= 100 %

AI FB1 damping

0 s

Damping

60 s

Real value correction

0

Reference value - Relay

None

Block adjustment

Released

Display

Language

Selected language

Displayed value

Pulse rate (ct/s)

With this function

Load parameter adjustment data from the sensor into the display

and adjustment module

Write parameter adjustment data from the display and adjustment

module into the sensor

This parameter is described in the operating instructions manual

"Display and adjustment module".
Info
In this menu you will find the following menu items:

Additional adjustments -

Copy instrument settings

Info