6 parameter adjustment - x-ray alarm – VEGA MINITRAC 31 Profibus PA User Manual

Page 61

Advertising
background image

61

6 Set up with the display and adjustment module

MINITRAC 31 • Profibus PA

43836-EN-131119

Load parameter adjustment data from the sensor into the display

and adjustment module

Write parameter adjustment data from the display and adjustment

module into the sensor

This parameter is described in the operating instructions manual

"Display and adjustment module".
Info
In this menu you will find the following menu items:

Instrument name - shows instrument name and serial number

Instrument version - shows hardware and software version of the

instrument

Date of manufacture - shows calibration date and the date of the

last change

Profibus Ident Number - shows the Profibus ident number

Instrument features - shows further instrument features

These parameters are described in the operating instructions manual

"Display and adjustment module".

6.6 Parameter adjustment - X-ray alarm

X-ray radiation, for example, cause with weld joint tests can influence

your measured values unnoticeably. Hence it is importnat to recog-

nized interfering X-ray radiation.
The MINITRAC 31 recognizes the X-ray raditation and reacts from an

individually definable raditiation volume.
The level or density measuring instrument acts as Master and the

MINITRAC 31 operates as Slave.
First of all, define the function of the Slave instrument before you

define the Master instrument. The Master instrument can thus imme-

diately recognize the connected Slave.
For this purpose, the Slave instrument must be defined as "X-ray

alarm". Select under the menu item "Setup/Application" the function

"X-ray alarm".
You can set the address (MGC) on the Slave instrument at will. Only

the address "0 - 0" is reserved for the Master instrument.
Set the address setting (MGC) on the Master instrument to "0 - 0".

You have to enter the address of the Slave instrument in the list of

the Master instrument. This function is not possible in the display and

adjustment module. To do this you need PACTware with the respective

DTM.

The main menu is divided into five sections with the following func-

tions:

Info

X-ray alarm

Main menu

Advertising