6 Set up with the display and adjustment module

MINITRAC 31 • Profibus PA

43836-EN-131119

You thus assign a corresponding density to the actual count rate.
Note:

If possible, take a product sample and determine the density simulta-

neously in your sampling position.
It has proven marking the product smaples with date and respec-

tive count rate. Hence the values can be assigned later on without

problems.

Accept the value pair with "OK".
Enter as many linearization points as possible. You can influence

the accuracy of the density measurement by doing this. The more

linearization points you enter and the higher the difference between

the density values, the more reliable the measurement.
Value pairs that are not yet complete, e.g. due to a missing density

determination, can be edited later with the function "Setup - Lineariza-

tion" under the item "Change/Edit".
A maximum of 32 linearisation points is possible.
Note:

If you cannot change the medium during the adjustment process, it

is possible to carry out the linearization with only one point. However,

you should enter further linearization points later, if possible.

Show diagram

This menu item is only available if a linearization was already carried

out.

Show table

In this menu item you can show the individual value pairs of the

linearization.

Linearization - Delete

You can also delete individual linearization points. Enter the number

of the point you want to delete.