VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) Profibus PA Approval according to LPR radio standard User Manual

Page 17

Advertising
background image

d

h

ma

x.

d

1½"
50 mm/2"
80 mm/3"
100 mm/4"
150 mm/6"

200 mm
250 mm
300 mm
500 mm
800 mm

h

max.

Fig. 9: Deviating socket dimensions

Tip:
The instrument is optionally available with antenna extension. The
antenna length can thus be selected so that the antenna end protrudes
at least 10 mm (0.4 in) out of the socket.

Align the sensor in liquids as vertical as possible to the product surface
to achieve optimum measurement results.

Fig. 10: Alignment in liquids

The mounting location of the radar sensor should be a place where no
other equipment or fixtures cross the path of the microwave signals.

Vessel installations, such as e.g. ladders, limit switches, heating
spirals, struts, etc., can cause false echoes and impair the useful echo.
Make sure when planning your measuring site that the radar sensor
has a "clear view" to the measured product.

In case of existing vessel installations, a false echo storage should be
carried out during setup.

If large vessel installations such as struts or supports cause false
echoes, these can be attenuated through supplementary measures.
Small, inclined sheet metal baffles above the installations scatter the
radar signals and prevent direct interfering reflections.

Sensor orientation

Vessel installations

VEGAPULS

62 • Approval according to LPR radio standard

17

4 Mounting

41720

-

EN

-120301

Advertising