VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) Profibus PA Approval according to LPR radio standard User Manual

Page 48

Advertising
background image

l

Basic settings: Resetting of the parameter settings incl. special
and laboratory parameters to the default values of the respective
instrument. Any created false signal suppression, user program-
mable linearization curve, measured value memory as well as
event memory will be deleted.

l

Setup: Resetting of the parameter settings to the default values of
the respective instrument . Order-related settings remain but are
not taken over into the current parameters. User-generated false
signal suppression, user-programmed linearization curve, meas-
ured value memory, echo curve memory as well as event memory
remain untouched. The linearization is set to linear.

l

False signal suppression: Deleting a previously created false
signal suppression. The false signal suppression created in the
factory remains active.

l

Peak values measured value: Resetting of the measured min.
and max. distances to the actual measured value.

Select the requested reset function [->] and confirm with [OK].

The following table shows the default values of VEGAPULS 62:

Menu

Menu item

Default value

Setup

Measurement
loop name

Sensor

Medium

Liquid/Water
Bulk solids/Crushed stones, gravel

Application

Storage tank
Silo

Vessel form

Vessel bottom, dished boiler end
Vessel top, dished boiler end

Vesell height/
Measuring range

Recommended measuring range, see
"

Technical data

"

in the supplement

Min. adjustment

Recommended measuring range, see
"

Technical data

"

in the supplement

Max. adjustment

0,000 m(d)

AI FB

1 Channnel PV (lin. perc.)

AI FB

1 scaling

unit

Height
%

AI FB

1 scaling

0.00 lin %, 0.00 %
100.00 lin %, 100.00 %

AI FB

1 damping

0 s

Lock operation

Released

Display

Language

Like order

Displayed value

SV

1

48

VEGAPULS

62 • Approval according to LPR radio standard

6 Set up with the indicating and adjustment module

41720

-EN

-120301