4 Mounting
VEGAFLEX 81 • HART - integrated accumulator
44640-EN-130430
1
2
Fig. 8: Installation in non-metal vessel
1 Flange
2 Metal sheet
If possible, avoid sockets. Mount the sensor flush with the vessel top.
If this is not possible, use short sockets with small diameter.
Higher sockets or sockets with a bigger diameter can generally be
used. They simply increase the upper blocking distance. Check if this
is relevant for your measurement.
In such cases, always carry out a false signal suppression after instal-
lation. You can find further information under "Setup procedure".
h
d
d
h
DN40 ... DN150
> DN150 ... DN200
150
100
_<
_<
Fig. 9: Mounting socket
When welding the socket, make sure that the socket is flush to the
vessel top.
Socket