21

5 Connecting to power supply

VEGAFLEX 81 • HART - integrated accumulator

44640-EN-130430

5 Connecting to power supply

5.1 Connection of the battery charger

We recommended charging the integrated accumulator completely

before setting up the instrument. You can find the charging time in

chapter "Technical data".
The battery charger is plugged into a socket in the supply room, see

chapter "Wiring plan".
The LEDs in the supply room show the charging process and condi-

tion of the accumulator, see chapter "Wiring plan".

If screened cable is necessary, connect the cable screen on both

ends to ground potential. In the sensor, the screen must be connected

directly to the internal ground terminal. The ground terminal on the

outside of the housing must be connected to the potential equalisa-

tion (low impedance).
If potential equalisation currents are expected, the connection on the

processing side must be made via a ceramic capacitor (e. g. 1 nF,

1500 V). The low-frequency potential equalisation currents are thus

suppressed, but the protective effect against high frequency interfer-

ence signals remains.

5.2 Wiring plan

2

1

Fig. 13: Position of the power supply and electronics compartment
1 Supply room (accumulator)

2 Electronics compartment

5 6 7 8

4...20mA

2

1

2

+

( )

(-)

1

Fig. 14: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment

2 Contact pins for the display and adjustment module

Cable screening and

grounding

Overview

Electronics compartment