4 Mounting

VEGAFLEX 81 • HART - integrated accumulator

44640-EN-130430

Note:

Measurement in a standpipe is not recommended for extremely

adhesive products.
Instructions for the measurement:

The 100 % point should not be above the upper tube connection

to the vessel

The 0 % point should not be below the lower tube connection to

the vessel

A false signal suppression with integrated sensor is generally

recommended to reach the maximum possible accuracy

If there is a danger of the cable probe touching the vessel wall during

operation due to product movements or agitators etc., the measuring

probe should be securely fixed.
In the gravity weight there is an internal thread (M8), e.g. for an eye-

bolt (optional) - (article no. 2.1512).
Make sure that the probe cable is not completely taut. Avoid tensile

loads on the cable.
Avoid undefined vessel connections, i.e. the connection must be

either grounded reliably or isolated reliably. Any undefined change of

this requirement can lead to measurement errors.

In case of difficult installation conditions, the probe can be also

mounted laterally. For this purpose, adapt the rod with rod extensions

or bow-shaped segments.
Let the probe length determine automatically by the instrument to

compensate the resulting running time changes.
The determine probe length can deviate from the actual probe length

when using bow-shaped segments.
Make sure that the rod of the probe is at least 300 mm (11.811 in)

away from the vessel wall
You can find further information in the supplementary instructions of

the rod extension.

Fixing

Rod extensions