VEGA VEGABAR 83 Profibus PA - Operating Instructions User Manual

Page 15

Advertising
background image

15

4 Mounting

VEGABAR 83 • Profibus PA

45037-EN-131107

1

2

3

4

4

4

Fig. 6: Position of the filter element - non-Ex, Ex-ia version
1 Housing plastic, stainless steel precision casting

2 Housing aluminium

3 Housing stainless steel, electropolished

4 Filter element

With the following instruments a blind plug is installed instead of the

filter element:

Instruments in protection IP 66/IP 68 (1 bar) - ventilation via capil-

laries in fix connected cable

Instruments with absolut pressure

The filter element is integrated in the process assembly. It is located in

a rotatable metal ring and has the following function:

Atmospheric pressure compensation (with relative pressure meas-

uring ranges)

Turn the metal ring in such a way that the filter element points down-

ward after installtion of the instrument. This provides better protection

against buildup.

2

1

3
4

3
4

Fig. 7: Position of the filter element - Ex-d version
1 Single chamber housing, aluminium, stainless steel precision casting

2 Double chamber housing, aluminium, stainless steel precision casting

3 Rotatable metal ring

4 Filter element

Instruments with absolute pressure have a blind plug mounted

instead of the filter element.

Instruments in Ex-d ver-

sion