VEGA VEGABAR 83 Profibus PA - Operating Instructions User Manual

Page 45

Advertising
background image

45

6 Set up with the display and adjustment module

VEGABAR 83 • Profibus PA

45037-EN-131107

Menu item

Parameter

Default value

Copy in-

strument

settings
Special pa-

rameters

No reset

Scaling

Scaling size

Volume in l

Scaling format

0 % corresponds to 0 l
100 % corresponds to 0 l

The instrument settings are copied with this function. The following

functions are available:

Read from sensor: Read data from sensor and store into the

display and adjustment module

Write into sensor: Store data from the display and adjustment

module back to the sensor

The following data or settings for adjustment of the display and ad-

justment module are saved:

All data of the menu "Setup" and "Display"

In the menu "Additional adjustments" the items "Reset, Date/Time"

The user-programmable linearization curve

The copied data are permanently saved in an EEPROM memory in

the display and adjustment module and remain there even in case of

power failure. From there, they can be written into one or more sen-

sors or kept as backup for a possible electronics exchange.
Note:

Before saving the data in the sensor, a safety check is made if the

data match with the sensor. For this purpose the sensor type of the

source data as well as the target sensor are displayed. If the data do

not match, a failure message is outputted or the function is blocked.

The data are only saved after release.

In this menu item you gain access to the protected area where

you can enter special parameters. In exceptional cases, individual

parameters can be modified in order to adapt the sensor to special

requirements.
Change the settings of the special parameters only after having con-

tacted our service staff.

Additional adjustments -

Copy instrument settings

Additional adjustments -

Special parameters