VEGA VEGABAR 83 Profibus PA - Operating Instructions User Manual

Page 9

Advertising
background image

9

3 Product description

VEGABAR 83 • Profibus PA

45037-EN-131107

Gauge pressure

Absolute pressure

Vacuum

Measured products are gases, vapours and liquids.
VEGABAR 83 is especially suitable for applications with higher tem-

peratures and high pressures.

The process pressure acts on the sensor element via the process

diaphragm and an internal transmission liquid. The process pressure

causes a resistance change which is converted into a corresponding

output signal and outputted as measured value.
Measuring ranges up to 16 bar: piezoresistive sensor element, meas-

uring ranges up to 25 bar: strain gauge (DMS) sensor element.

1

3

4

2

Fig. 3: Configuration of the measuring system with piezoresistive sensor ele-

ment
1 Sensor element

2 Base element

3 Transmission liquid

4 Process diaphragm

With small measuring ranges or higher temperatures, the METEC

®

measuring cell is the measuring unit. It consists of the ceramic-capac-

itive CERTEC

®

measuring cell and a special, temperature-compen-

sated chemical seal system.

Measured products

Measuring system