VEGA VEGABAR 83 Profibus PA - Operating Instructions User Manual

Page 56

Advertising
background image

56

9 Diagnosis, asset management and service

VEGABAR 83 • Profibus PA

45037-EN-131107

9.5 Calculation of total deviation (according to

DIN 16086)

The total deviation of a pressure transmitter indicates the maximum

measurement error to be expected in practice. It is also called maxi-

mum practical deviation or operational error.
According to DIN 16086, the total deviation F

total

is the sum of the

basic accuracy F

perf

and the long-term stability F

stab

:

F

total

= F

perf

+ F

stab

The basic accuracy F

perf

is composed of the deviation F

Kl

and the

thermal change of zero signal and output span F

T

:

F

perf

= √((F

T

)

2

+ (F

Kl

)

2

)

This applies to the digital signal output via HART, Profibus PA or

Foundation Fieldbus. With a 4 … 20 mA output, there is also the

thermal change of the current output F

a

:

F

perf

= √((F

T

)

2

+ (F

Kl

)

2

+ (F

a

)

2

)

To provide a better overview, the formula symbols are listed together

below:

F

total

: Total deviation

F

perf

: Basic accuracy

F

stab

: Long-term drift

F

T

: Thermal change of zero signal and output span

F

Kl

: Deviation

F

a

: Thermal change of the current output

You can find the specifications in chapter "Technical data".

Pressure measurement in the pipeline 8 bar (800 KPa)
Product temperature up to 50 °C
VEGABAR 83 with measuring range 10 bar, deviation < 0.075 %
Calculation of the turn down: TD = 10 bar/8 bar, TD = 1.25
1. step: Calculation of the basic accuracy

F

perf

= √((F

T

)

2

+ (F

Kl

)

2

)

F

T

= (0.15 %) x (0.5 + 0.5 x TD)

F

Kl

= 0.075 %

F

perf

= √((0.15 %) x (0.5 + 0.5 x TD)

2

+ (0.075 %)

2

)

F

perf

= 0.19 %

Step 2: Calculation of the total deviation

F

total

= F

perf

+ F

stab

F

stab

= (0.05 % x TD)

F

stab

= (0.05 % x 1.25)

F

stab

= 0.06 %

F

total

= 0.19 % + 0.063 % = 0.25 %

1. step: Calculation of the basic accuracy

F

perf

= √((F

T

)

2

+ (F

Kl

)

2

+ (F

a

)

2

)

Meaning of the total

deviation

Practical example

Calculation - Digital out-

put signal

Calculation - Analogue

output signal

Advertising