FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER

2

SURETEST BASE UNIT

Rear Panel (Input/Output)




Figure 2-3 SURETEST Rear Panel

A -

Solenoid Control

(Hirose RM15TRD-12S)

(for use with optional 2000-600-02 Loaders)
PIN FUNCTION
10

Loader Relay Common

11

Loader Relay CCW Limit

12

Loader Relay CW Limit


B -

Analog Output (BNC)

Analog output provides a voltage level output that is directly proportional to transducer input,
from -1.8V (full scale counter clockwise) to +1.8V (full scale clockwise). Zero offset +/-100 mV.
Linearity +/-1% of reading. Maximum load = 5 mA.

When using the 4-in-1 transducer, zero output fluctuates until one of the transducers is scan
selected by applying 5% of full scale torque.

C -

Foot Pedal Switch Input

(used to send output to printer)


D -

Fuse Holder

Use AC fuse at specified rating only. ( 3.15 Amp).



E -

Input AC Voltage
Automatic Selection of any AC voltage from 100VAC to 230VAC, 50-60 Hz, 50 W.