CALIBRATION

CHAPTER 5

13. Press (Accept) to accept the calibration value (1

st

Cal point).

Apply 20% Torque = 10 in. lb.

14. Press

(Accept) to accept the calibration value (2

nd

Cal point) .

Apply 40% Torque = 20 in. lb.

16.

Press (Accept) to accept the calibration value (3

rd

Cal point) .

Apply 60% Torque = 30 in. lb.


17. Press (Accept) to accept the calibration value (4

th

Cal point) .

Apply 100% Torque = 50 in. lb.

18. Press (Accept) to accept the calibration value (5

th

Cal point) .

The display will show: Saving …. then show the Calibration table.

19. Remove all weights and the arm.

Press any key, go back to Select mode.

Calibration CCW:

(Go to step 7 then select CCW)

20. Select "Calibrate CCW” on the LCD display.

Apply and release full scale torque to the transducer three times in the CCW direction then
remove the weights and hanger, wait for about 15 seconds.

21. Repeat steps 12 through 19, applying all torque in the counter clockwise (CCW) direction.