CALIBRATION

CHAPTER 5

The 4-in-1 transducer may also be mounted to the calibration stand, shown in Figure 5-1, using the
standoff fixture and bracket assembly. For this mounting configuration the two slide knobs must be
loosened and four 5/16-18 bolts attached to secure the 4-in-1 transducer to the bracket assembly.

Quick Check

WARNING

Always be alert to the potential for personal injury that may be caused by excessive

torque applications, careless handling of heavy weights, and out-of-balance or unsafe

weight distribution.

1. With

the

SURETEST ON, select TRACK mode, and select the UNITs of measure appropriate to

the transducer as specified in the calibration checkpoint table in this chapter.

2.

Install the transducer to be checked and allow a 5-minute warm up period to stabilize the electronic
circuits and transducer elements.

3.

Transducers are mounted horizontally for calibration as instructed in Mounting Details in this
chapter. Mount the 4-in-1 transducer to a sturdy bench using the bracket provided. Mount single
transducers horizontally using the accessory stand and block. Attach the specified torque test bar,
wheel or butterfly, secured with its drive retention screw, and install the specified hanger for
clockwise (CW) torque application.


4.

Apply and remove the full range torque for the transducer three times.


5.

Remove the weights and hanger, but leave the arm attached, and press Zero (F1).

6.

At each of the check points specified for the transducer in the calibration check point table (in this
chapter), apply torque in the CW direction and verify the display readings are within 0.25% of
applied torque. If any reading is out of specification, proceed to Calibrating SURETEST Torque
Transducers in this chapter.


7.

Repeat Steps 3 through 6 in the counter clockwise (CCW) direction.

When using the 2-1/2" or 5" test wheel, a small “S” hook is supplied to provide a positive

zero bias in the direction (CW or CCW) that the calibration is performed. Do not remove

the “S” hook when setting zero tare.