CALIBRATION

CHAPTER 5

Press

Unit (F2) key and select the desired torque unit.

10.

Using the Up (

↑) and Down (↓) Cursor keys, enter the Full Scale value for the 4-in-1 transducer to calibrate.

Possible values are:
1

=

50 in. lb.

2

=

400 in. lb.

3

=

1000 in. lb.

4

=

250 ft. lb.

( For 4 in 1 Transducers display shows : CHECKING TRANSDUCER.)
Apply torque to one of the transducers until “ Full Scale UNITS shows on display.

Note:

Example graphics are for 50 in-lb transducer.

11.

Press

ENTER (

↵).



12.

Press any key to initiate Zerotare. The LCD displays:

then:

13.

Apply torque using the precision arm, hanger or certified weights listed as calibration point 1 in the
calibration check point table in this chapter. When the reading is stable, push and hold the Up (

↑) or Down

(

↓) Cursor Keys to get the correct torque value reading. Repeat the procedure for points 2, 3, 4, and 5 of the

same table as shown below:

Apply 10% Torque = 5 in.lb

5-9