Xylem ProLab 3000 LABORATORY pH METER User Manual

Page 103

Advertising
background image

ProLab 3000

Ion concentration

103

ba75645e02

01/2010

Calibration evaluation

After calibrating, the meter automatically evaluates the calibration.

Display

Calibration
record

Magnitude of the slope [mV]

Ions with
valency 1

Ions with
valency 2

+++

50.0 ... 70.0

25.0 ... 35.0

Error

Error

< 50 or > 70

< 25 or > 35

Eliminate the error according
to chapter 15 W

HAT

TO

DO

IF

...

(page 165)