System functions

65

ba75645e02

01/2010

6.10

Reset

You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.

6.10.1 Resetting the sensor settings

The sensor settings are reset in the sensor menu.

In the measured value display, activate the relevant sensor window
[Tab] and open the menu with [Menu].

pH

The following settings for pH measurements are reset to the default
settings with the Reset pHMeasuring function:

6.10.2 Resetting the system settings

The following system settings are reset to the default settings with the
System / Reset function:

Note
The calibration data is reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.

Setting

Default settings

Calibration interval...

7 d

Slope in

mV/pH

Measured parameter

pH

pH resolution

High

Zero point

0 mV

Slope

-59.16 mV/pH

Temperature, manual

25 °C

Setting

Default settings

Language

English

Temperature unit

Celsius

Signal On

Baud rate (Baud)

4800

Output format

ASCII