1 rs232 interface (serial port), Rs232 interface (serial port) – Xylem ProLab 3000 LABORATORY pH METER User Manual

Page 29

Advertising
background image

ProLab 3000

Commissioning

29

ba75645e02

01/2010

3.7.1

RS232 interface (serial port)

PC and printer can be connected via the RS232 interface.

Data that is transmitted to the interface with <PRINT> can be printed
out with a printer or received with a terminal program if a PC is con-
nected.

Operation with RS232

Connect the interface of the ProLab 3000 to the devices using the Z390
(PC) or Z893 (external USB printer Z890) cable.

Set up the following transmission data on the PC/printer:

Baud rate

selectable between: 1200 ... 115200

The baud rate must agree with the baud rate set on
the PC/printer.

Handshake

RTS/CTS

PC only:

Parity

none

Data bits

8

Stop bits

1

Note
Pin assignment of the socket: See page 169.