10.2.2 Managing the calibration data

The calibration data is stored in the current calibration record (menu,
Calibration / Calibration record...) and in the calibration history (menu,
Calibration / Calibration history... and Memory / Calibration memory).

The current calibration record of ID sensors is stored in the sensor, of
non ID sensors it is stored in the ProLab 3000.

The meter creates a separate calibration history for each ID sensor cal-
ibrated with the ProLab 3000.

For non ID sensors calibrated with the meter, a calibration history is
created for each measured parameter (pH, ISE) where the current cal-
ibration data is appended.

Managing functions for all calibration data are available to the adminis-
trator only.

Viewing the calibration

history

The administrator can view the calibration histories of all sensors and
users.

Erasing a calibration

history

The administrator can delete the calibration histories of all sensors and
users.

The current calibration data of each sensor is stored separately and is
retained even if the calibration history is deleted.

1

Open the menu, Memory / Calibration memory / Display....
The Open file dialog box pops up.

2

If necessary, select a different folder.

3

Select a file ("*.cal") and confirm with [OK].
The Calibration memory window with the calibration data pops
up.

4

If necessary, output the calibration data to the interface with
[Print].

1

Open the menu, Memory / Calibration memory / Delete....
The Delete file dialog box pops up.

2

If necessary, select a different folder.

3

Select a file ("*.cal") and confirm with [OK].
A safely query pops up.

4

Confirm the deletion of the file with [OK].
The file is deleted.