2 calibration data, 1 displaying stored calibration data – Xylem ProLab 3000 LABORATORY pH METER User Manual

Page 136

Advertising
background image

Memory

ProLab 3000

136

ba75645e02

01/2010

10.2

Calibration data

The current calibration record is automatically stored after each valid
calibration procedure.

10.2.1 Displaying stored calibration data

Displaying the current

calibration record

The current calibration record for a meter and a sensor can be found in
the menu, Calibration / Calibration record....

Displaying the

calibration history

1

Highlight a channel in the measured value display.

2

Select the measured parameter with <MODE>.

3

Open the Calibration menu with Calibration record....
The current calibration record is displayed.

4

If required, output the calibration data to the interface with
[Print].

Note
The current calibration record can be quickly displayed by clicking on
the CalClock with the right mouse button.

1

Open the Calibration / Calibration history... menu.
The Calibration memory window with the calibration data for
the selected sensor pops up.

2

If required, output the calibration data to the interface with
[Print].