ProLab 3000

Overview

19

ba75645e02

01/2010

1.5.2

Sensor data from ID sensors

ID sensors transmit the following sensor data to the meter:

z SENSOR ID

– Sensor type

– Sensor series number

z Calibration data

– Calibration date

– User (having carried out the last calibration procedure)

– Calibration characteristics

– Calibration interval

– Selected buffer set (pH electrodes only)

The calibration data is updated in the ID sensor after each calibration
procedure. A message is displayed while data is updated in the sensor.

Note
While data is updated in the sensor, the sensor must not be discon-
nected because otherwise the calibration data will not be completely
transmitted. The sensor will then have no valid calibration.

Note
If non-ID sensors are used, the calibration data from the meter is used
and also stored in the meter.