8 automatic stability control, Automatic stability control – Xylem ProLab 3000 LABORATORY pH METER User Manual

Page 63

Advertising
background image

ProLab 3000

System functions

63

ba75645e02

01/2010

6.8

Automatic stability control

The Autom. stability control function continuously checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values.

You can activate or switch off the Autom. stability control function
(menu, Autom. stability controlSystem).

The Autom. stability control function is carried out:

z as soon as the measured value is outside the allowed stability range

z when you switch over between the measured parameters with

<MODE>.

Whether or not a measured value is stable can be taken from its color
labeling.

In the delivery condition, the SC stable / SC not stable conditions are
assigned to the following colors:

Color

Meaning

red

SC not stable
Stability criteria not met

green

SC stable
Stability criteria met

SC stable

SC not stable