Ion concentration

ProLab 3000

124

ba75645e02

01/2010

6

In the Standard register, enter the values for Standard volume-
Standard concentration and ISA/TISAB volume if necessary.

7

Check the currently displayed ions.
If the ion of the standard (e.g. Cl (Sub) is displayed incorrectly:

Change the ion type of the standard solution [e.g. Cl (Sub)] in
the ION register.

8

Thoroughly rinse the electrode with deionized water.

9

Immerse the electrode in the sample.

10

Start the measurement with [Continue] .

The measured value is checked for stability (Stability control).

11

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Sample register opens up.

12

Add the sample to the standard solution.

13

In the Sample register, enter the value for the sample volume
(Sample volume) if necessary.

14

Start the measurement with [Continue].
The measured value is checked for stability (Stability control).

15

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.