Factory Restore Command
FACTORY_RESTORE parameter
This parameter is a write only command that restores the instrument calibration to the calibration parameters that were set by the factory,
for the presently active calibration group.
Individual Sensors Data View
This section of the Service Transducer Block is a read only section. It shows a snap shot of the process data that each individual sensor
head is detecting. This section provides information for up to 16 sensors of a multi-point ST100 instrument.
Typical Parameters for a Sensor Head (Shown is Sensor 1)
FLOW_VALUE_SENSOR_1 parameter
This parameter presents the flow value of flow sensor 1 in SFPS.
TEMPERATURE_VALUE_SENSOR_1 parameter
This parameter presents the temperature value associated with flow sensor 1 in °F.
PRESSURE_VALUE_SENSOR_1 parameter
This parameter presents the pressure value associated with flow sensor 1 in PSIA.
ST100 Service Transducer Block
The table below summarizes the ST100 Service Transducer Block.
BLOCK INFO
INDEX
PARAMETER
DATA TYPE (LENGTH)
INITIAL VALUE
DESCRIPTION
0
BLOCK OBJECT
DS-64
1
ST_REV
Unsigned16
0
2
TAG_DESC
Octet String
spaces
3
STRATEGY
Unsigned16
0
4
ALERT_KEY
Unsigned8
0
5
MODE_BLK
DS-69
0
6
BLOCK_ERR
Bit String
0
7
UPDATE_EVT
DS-73
8
BLOCK_ALM
DS-72
9
TRANSDUCER_DIRECTORY
Unsigned16
0
10
TRANSDUCER_TYPE
Unsigned16
65534
11
XD_ERROR
Unsigned8
0
12
COLLECTION_DIRECTORY
Unsigned32