3.3.1 Installation in Conduit

In hard-wired electrical applications, the power and safety ground
service drops for the instrument must be 18 to 12 AWG. See

Figure 10

for strain relief and conduit opening sealing plug

information. See

section 3.3.3 on page 14

for wiring information.

3.3.2 Installation Using a Power Cord

DANGER
Use of a power cord is not acceptable in Class 1, Division 2
Hazardous Location Installation (see).

A sealing-type strain relief to maintain the NEMA 4X/IP66
environmental rating and a power cord less than 3 meters (10 feet)
in length with three 18-gauge conductors (including a safety ground
wire) can be used, see

Replacement Parts and Accessories on

page 41

. See

Figure 10

for strain relief and conduit opening sealing

plug assembly. See

section 3.3.3 on page 14

for wiring information.

Figure 9 Removing Voltage Barrier

1

High voltage barrier

2

Unsnap the barrier latch then pull out to remove the barrier.

NC

NC

NC

COM

COM

COM

NO

F1

F2

NO

NO

RELAY C

RELAY B

RELAY A

1

1

+

DATA

+

OUT 2

!

DATA

!

OUT 2

SER

VICE REQUEST

SHIELD/CHASSIS GND

+

V

+

OUT 1

GND

! OUT 1

2

2

3

3

4

4

5

5

6

PROBES

ANAL

OG

OUTP

UTS

PCB

CONNECTOR

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

DANGER - EXPLOSION HAZARD

DANGER - RISQUE D'EXPLOSION

DO NOT DISCONNECT WHILE CIRCUIT IS LIVE

UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.

NE PAS DEBRANCHER TANT QUE LE EST SOUS

TENSION, A MONIS QU'IL NE S'AGISSE D'UN

EMPLACEMENT NON-DANGEROUX

J1

J2

J4

NETWORK

INTERFACE

CARD

J3

J5

J6

U5

U9

S1

2

1

Figure 10 Using the Optional Strain Relief and Conduit Plug

1

Power cord strain relief

2

Conduit strain relief

3

Conduit opening sealing plug

2