Installation – Hach-Lange SC 60 User Manual

Page 26

Advertising
background image

22

Installation

Figure 18

Quick-connect Fitting Pin Assignment

3.5.1.2 Hard-wiring a sc Sensor to the Controller

1.

Open the controller cover.

2.

Disconnect and remove the existing wires between the quick
connect and terminal block J5, see

Figure 19 on page 23

.

3.

Remove the quick connect fitting and wires and install the
threaded plug on the opening to maintain the environmental
rating.

4.

Strip the insulation on the cable back 1-inch. Strip ¼-inch of
each individual wire end.

5.

Pass the cable through conduit and a conduit hub or a strain
relief fitting and an available access hole in the controller
enclosure. Tighten the fitting.

6.

Wire as shown in

Table 6: on page 24

.

7.

Close and secure the cover.

Item

Designation

Wire Color

1

Pin 1 (+12 V DC)

Brown

2

Pin 2 (Circuit Common)

Black

3

Pin 3 (Data (+))

Blue

4

Pin 4 (Data (–))

White

5

Pin 5 (Shield)

Shield (grey wire in existing quick-disconnect fitting)

6

Groove

1

2

3

4

5

6