19

Installation

3.4.2 Connecting the Analog Outputs

DANGER
Explosion hazard. Do not connect or disconnect electrical
components or circuits to the equipment unless power
has been switched off or the area is known to be
non-hazardous.

Two isolated analog outputs (1 and 2) are provided, see

Figure 16

on page 20

. Each output can be set to 0–20 or 4–20 mA, and can

be assigned to represent the measured parameter or secondary
measurement such as temperature. Make connections with
twisted-pair shielded wire and connect the shield at the controlled
component end or at the control loop end. Do not connect the
shield at both ends of the cable.
Use of non-shielded cable may
result in radio frequency emission or susceptibility levels higher
than allowed. Maximum loop resistance is 500 Ohm. Refer to the
sensor manual for output software setup.

Make wiring connections at the analyzer end as shown in

Table 4:

and

Figure 16 on page 20

.

Figure 15 Alarm and Relay Connections

NC

NC

NC

COM

COM

COM

NO

F1

F2

NO

NO

RELAYC

RELAYB

RELAYA

J1

J2

J4

S1

NETWORK

INTERFACE

CARD

J3

J5

J6

U5

U9

J5

J5

NC

NC

NC

COM

COM

COM

NO

F1

F2

NO

NO

RELAYC

RELAYB

RELAYA

J1

J2

J4

S1

NETWORK

INTERFACE

CARD

J3

J5

J6

U5

U9

J5

J5

J5

J6

J7

3

2

1

RELAYA

RELAYA

RELAYB

RELAYB

RELAYC

RELAYC

Disconnect

Power

NC

COM

NO

Advertising